Accurate Continuous Monitoring of Oxygen in Hydrogen / Nitrogen Atmospheres used in Annealing Furnaces
The Novatech 1738 detects changes in the dew point of annealing furnaces allowing operators to take corrective action to prevent expensive re-work due to oxidation of the product. The dew point is calculated from the amount of oxygen measured in the furnace. The unique sensor manufactured by Novatech uses a catalyst that attracts oxygen molecules in the same way as the surface of the steel does inside the furnace.
Using the Novatech 1738 is easy and convenient
- Simply enter the percentage of hydrogen used by the furnace.
- Very low maintenance and re-calibration
- Two fully programmable isolated 4-20mA outputs
- Built in Modbus™ slave protocol and RS-485 networking
Dual sensors provides redundancy to prevent costly shutdowns if a probe failure occurs
The Novatech 1738 Dew Point Transmitter accepts inputs from two sensors, averaging the two readings, or providing separate 4-20mA outputs for each sensor. In case of either one of the sensors failing, the transmitter warns the operator and locks onto the remaining sensor so that the process can continue to operate.
The Novatech 1738 tells you what’s happening
Plant operators are alerted to failure of the probe by a plain English message on the transmitter’s LCD display. There are 30 probe and transmitter alarms, plus 4 process alarms. The upper line display indicates: Oxygen %. Auto-ranging from 1 x 10-30% to 100% oxygen.
Specifications
| Inputs |
|---|
- One or two zirconia oxygen probes or sensors
- One zirconia sensor & auxiliary thermocouple type J, K, R or S
- Burner “On” signal (dry contact)
- Purge air flow switch
|
| Outputs |
|---|
- Four programmable alarm relays
- Two isolated 4-20mA or 0-20mA
- SSR outputs to purge & calibration check gas solenoid valve
|
| Range of Outputs |
|---|
- Dew Point-60 to 40°C
- Average Dew Point-60 to 40°C
- Linear Oxygen0 to 100%
- Average Linear Oxygen0 to 100%
- Reducing Oxygen10-30 to 102%
- Average Reducing Oxygen10-30 to 102%
- Pre-Reactive Oxygen0 to 10%
- Average Pre-Reactive Oxygen0 to 10%
- Probe EMF0 to 1300mV
- Auxiliary TC Temperature0 to 1400°C
|
| Alarms |
|---|
- Common alarm relay with 20 user selectable instrument alarm functions.
- Three programmable process alarm relays for
- Oxygen High
- Oxygen Deviation
- Dew Point High
- Dew Point Deviation
- Pre-Reactive Oxygen High (two separate thresholds)
- Probe Temperature Low
- Cal. Check in Progress
- Purge in Progress
- Any alarm condition not selected for the common alarm
- Multiple selections can be made for all relays
|
| Alarm Contacts |
|---|
- Normally open failsafe (open for alarm state)
- 250VAC / 30VDC, 2A
|
| Range of Local Indication |
|---|
- 1.0 x 10-30% to 100% Oxygen
- 0.01ppm to 10,000ppm – automatically defaults to exponential format below 0.01ppm and percent format above 10,000ppm (1%)
|
|
| Network Interface |
|---|
|
|
| Secondary Parameter Display |
|---|
- Probe 1/2 Dew Point
- Average Dew Point
- Probe 1/2 TC Temperature
- Probe 1/2 EMF
- Probe 1/2 Impedance
- Probe 1/2 Oxygen %
- Average Oxygen %
- Probe 1/2 Pre-Reactive O2
- Average Pre-Reactive O2
- Auxiliary TC Temperature
- Ambient Temperature
- Ambient RH%
- Runtime
- Service Date
- 4-20mA Output 1
- 4-20mA Output 2
Probe 2 & Average Parameters only available with two probes |
| Accuracy |
|---|
- ± 1% of the actual oxygen reading with a repeatability of 0.5%. For example, at 2% oxygen the accuracy would be ±0.02% oxygen.
|
| Environmental Rating |
|---|
- Operating Temperature -25°C to 55°C
- Relative Humidity 5% to 95% (non-condensing)
- Altitude 2000m Maximum
|
| Power Requirements |
|---|
- Mains Voltage 100 to 240VAC -6 +10%, 50/60Hz
- Overvoltage Category II (IEC60364-4-443)
- Requires 5W for controller plus probe power
|
| Degree of Protection |
|---|
- IP65
- IP54 with internal reference air pump
|
| Dimensions |
|---|
- 315mm W x 190mm H x 110mm D
|
| Weight |
|---|
|
|
|