Oxygen Measurement Made Simple

The Novatech 1234 Oxygen Sensors are ideal for:

Reasons for using Novatech 1234 sensors

The Novatech 1234 Oxygen Sensors

Accuracy and reliability

The Novatech 1234 Oxygen sensors provide accurate and virtually drift-free measurement for years.

Adapting the Novatech 1234 to your application

The Novatech 1234 sensors are available for Oxygen measurement in a range of applications by using ancillary equipment including:

The Novatech 1234 sensors are responsive

Accurate readings can be made with sample response times as short as 3 seconds to changes in gas composition for Oxygen.

Specifications

Measuring Range
  • 1 ppm to 100% oxygen
Sensor Output
  • EMF = 2.154 x 10-2 x T x loge(0.209 / oxygen level)
Response Time
1234C 5 seconds with a gas flow of 2 litres / min
1234E 3 seconds with flows from 0.5 to 25 litres / min
1234M 3 secs with flows from 0.1 to 0.5 litres / min
Accuracy
  • ±1% of actual reading
Thermocouple
  • Type K
Warm Up Time
  • Seven to ten minutes
Heater Power
  • 110 VAC, 50 / 60 Hz, 115Watts
Flow Rate Range
  • 0.1 to 25 litres / min
Gas Connection
  • 1/4” NPT female, inlet and outlet
Sample Aspiration
  • Convection flow from hot furnace or flue, from process pressure, electric pump or air operated ejector (for 1234E model only).
Environmental
  • Non weather proof cabinet. If mounting outdoors, a weather-hood is required for electrical protection. Do not enclose in sealed cabinet or overheating may occur. Your cabinet should be vented.
  • Ambient Temperature 0 - 100°C
Weight
  • 2.2 Kg
Dimensions
  • 300 mm x 125 mm x 88 mm
Optional Sample Pump
  • External electrical diaphragm pump, 240 / 110 VAC, or air operated ejector, 30 to 100 kPa
Optional Process Sampling Probe, Filter and Bush
  • Stainless steel filter, 500ºC max. Process connection bush 1.5” BSP or NPT to 3/8” tube. (3/8”, 16 gauge tube supplied by others)
Mounting
  • Vertical surface mount, brackets supplied
Ordering Information
1234C If sample is delivered from process or pump
1234E If sample is drawn from the outlet by suction, specify mains voltage on 1234E, 240V or 110V
1234M For low flow rate requirement
Accessories
  • Integral electric pump and filter assembly MV-10F (fitted on top of your 1234 sensor)
  • Air operated ejector (for 1234E model only)
  • Sampling probe filter and bush, tube by others
  • Filters for dry dust, wet dust or hydrocarbons
  • Flow meter